AlphaScreen SureFire Bead Titration Protocol
top
Below is an example of how to make your master mixes to perform a bead titration for most 2-step kits (where the original protocol recommends mixing the Reaction buffer, Activation buffer, and Acceptor beads in a 40:10:1 ratio). This ratio is indicated in the manual, in the "Buffer preparations and Storage Conditions" table. For other kits, you can titrate the beads similarly by reducing only the volume of beads in the mix (by a factor of 2 or 4).
Though the final volume of the mixes (and therefore the concentrations of reaction buffer, activation buffer, dilution buffer, and Donor and Acceptor beads) changes slightly, we recommend that you perform the bead titration experiments by altering the volume of the master mixes in this manner.
- Application Support Knowledge Base Home
- Alpha SureFire No-wash Cellular Kinase Assays
- Alpha SureFire Assay Development
- Alpha SureFire Tips and FAQs
- AlphaScreen SureFire Protocol Adaptations
- AlphaScreen SureFire Bead Mix Volumes
- AlphaScreen SureFire Bead Titration Protocol
- AlphaScreen SureFire Troubleshooting
- AlphaScreen SureFire Citations
- AlphaScreen SureFire Instrument Options and Settings