Radiometric Ligand-Binding Assays

Overview


Radioactive ligands are commonly used to measure ligand binding to receptors. In this assay, you will measure binding of a radiolabeled ligand to cells or cell membranes containing a receptor of interest. Both whole cells and cell membranes can be used for this assay. Radioligands can be used to perform saturation curves, competition and kinetic experiments.

radioligand_main_principle_ASK.jpg
Ligand-receptor binding

When selecting a radioligand, there are a few factors to keep in mind:

  • High specific activity: Radioligands with a high specific activity are well-suited for ligand binding assays. Specific activity indicates how much radioactivity there is per molecule of ligand, and is usually given in units of Curies per millimole of ligand. Several of our 125I-labeled ligands are offered at maximum specific activity (2200 Ci/mmol if one 125I labeling site is available, 4400 Ci/mmol if two 125I labeling sites are available, etc.). This indicates that virtually every molecule of ligand provided in the stock vial is radiolabeled. For tritiated ligands (3H ligands), you should ideally pick a ligand that has a specific activity above 20 Ci/mmol. The maximum theoretical specific activity per tritium is 29 Ci/mmol (Curies per millimole of tritium). Specific activities above this value indicate that on average, each molecule of ligand has at least one tritium.
  • Low non-specific binding: Hydrophobic ligands will generally show higher non-specific binding. This can sometimes be compensated by coating filters with BSA to reduce non-specific binding. Including BSA, salts or detergents in the wash or binding buffer can also help reduce non-specific binding. If the stock radiochemical is packaged in a silanized vial (refer to the tech data sheet), this may indicate the ligand is somewhat hydrophobic
  • High purity: Ideally, the ligand should have a radiochemical purity above 90%. Radiochemical purity decreases over time, and the actual rate of this degradation accelerates over time. Radiochemical purity and degradation rates for our various radiochemicals can be found on the lot-specific technical data sheets.
  • High selectivity: The more selective the ligand is for your receptor, the better your data will be. High selectivity indicates the ligand will mostly be recognized by only one receptor-of-interest, over other receptors that may be present in a cell membrane. Also, using membranes that over-express the receptor of interest may help reduce any potential contribution from receptors endogenously expressed in the membrane.
  • Stability: If you will need to use your radioligand over an extended period of time, stability may be a factor for you. 125I-labeled ligands should generally be used within one to two months of the manufacture date. Tritiated ligands should usually be used within 3-6 months of manufacture date; however, there are exceptions to this. Degradation rates and manufacturing dates can be found on our lot-specific technical data sheets. You can also contact technical support to discuss the recommended use time for each 爱游戏平台注册登录 radiochemical - our contact information is on the upper right-hand corner of this page.
  • Energy3H releases beta energy, which can be measured on a scintillation counter after the addition of scintillant (in the form of a scintillation cocktail, Meltilex®, an SPA bead, a FlashPlate®, etc.). The beta energy interacts with the scintillant to produce photons, which are measured by the detector. 125I releases both beta-like energy and gamma energy. If you only have access to a gamma counter, you should use a radioligand labeled with 125I. An additional factor is assay format. 125I gives off more beta-like energy than 3H. If you will be performing a FlashPlate assay, it will be easier to develop an assay using a radioligand labeled with 125I because the scintillant is embedded within the walls of the plate (greater distance between radioligand and scintillant). For SPA assays, either 3H or 125I-labeled ligands can be used effectively.

Radioactive ligand binding assays can be performed is several different formats. Typically, we perform this assay in filtration format. The assay can also be performed in SPA formats, or in FlashPlates®. SPA and FlashPlate formats are homogeneous formats, which means no wash steps are required. In the filtration assay, you will wash away unbound ligand using a vacuum manifold or cell harvester.

 SPA AssayFlashPlate AssayFiltration Assay
FormatHomogeneousHomogeneousWash-based
Advantages
  • No wash steps required
  • Can be run in 96-well, 384-well formats
  • Assays can usually be optimized within one day
  • Typically generates higher Z’ values due to lower assay variability
  • No wash steps required (though you could wash if desired)
  • No potential for bead settling
  • Can be run in 96-well, 384-well formats
  • Typically generates higher Z’ values due to lower assay variability
  • Assays can usually be optimized within one day
  • Separation technology: can have a better window in some cases
  • Can work with larger volumes
  • Fewest assay development steps
Disadvantages
  • Background can sometimes be an issue, which may require further assay optimization
  • You will need to carefully titrate the amount of cell membrane when you are optimizing your assay
  • Background can sometimes be an issue, requiring further assay development
  • Because this is a wash-based assay, you might have more assay variability. You should carefully optimize wash steps/volumes
  • More radioactive waste is generated
  • Cannot be run in high-throughput (the filtration assay is usually run in 96-well format)
Recommended forLow-throughput (just a few assays) or high throughput; automation; reducing amount of radioactive waste generatedLow-throughput (just a few assays) or high throughput; automation; reducing amount of radioactive waste generatedLow-throughput assays or just a few experiments; may be desirable if your lab is already set up for radioactive filtration assays

Top

Assay Formats


Filtration Ligand Binding Assays

View more information on what materials you need, what optimizations you may need to perform, and references for radioactive ligand binding filter plate assays. In filtration format, the binding assay is carried out first in one assay plate, then filtered through a filtermat or UniFilter® plate using a cell harvester (vacuum manifold). The filters are washed to remove any unbound radioligand. The filtermat or UniFilter plate is then dried and scintillation cocktail (or Meltilex®) is added before reading in an appropriate detector.

radioligand_filtration_ASK.jpg

SPA ligand binding assays

View more information on what materials you need, what optimizations you may need to perform, and references for SPA radioligand binding assays. In the SPA format, cell membranes are captured onto SPA beads. When radioligand binds to the receptor/membrane, this puts the radiochemical into proximity of the SPA bead. The beta energy from the radioligand can interact with scintillant in the bead, producing a signal that can be measured. Radioligand that is not bound to the cell membrane will not be close enough to the SPA bead to interact strongly with the scintillant.

radioligand_spa_ASK.jpg

FlashPlate ligand binding assays

View more information on FlashPlate ligand binding assays. The binding characteristics of FlashPlates are similar to other polystyrene microplates, allowing the use of standard published plate-coating methods. A target receptor or receptor membrane is coated or bound onto the wall of a FlashPlate well, after which radioligands, standards, and compounds are added. Energy emitted from the radioligand can interact with scintillant embedded in the walls of the FlashPlate, producing a detectable signal. Unbound radioligand is not close enough to the walls of the plate to activate the scintillant, which allows you to distinguish radioligand bound to the receptor/membrane from radioligand free in solution. No separation step is required.

radioligand_flashplagte_ASK.jpg

Top

Overview of ligand binding


Definitions

  • Affinity (potency): the tightness with which the ligand binds to the receptor. This is usually expressed as an equilibrium constant, Kd. The lower the Kd value, the higher the affinity.
  • Specificity: describes how selective a ligand is for one receptor vs. other similar receptors
  • Kd: The concentration where 50% of the receptors are occupied by radioligand

Pharmacological Classifications for Ligands

partial_agonist_ASK.jpg
Ligands

  • Full agonist: a ligand that binds to and activates a receptor and triggers a maximal response
  • Partial agonist: a ligand that binds to a receptor and produces a sub-maximal response
  • Antagonist: a ligand that binds to a receptor that does not activate the receptor.  Antagonists trigger no response when applied on their own.  They will block the effects of a competing agonist ligand.
  • Inverse agonist: a ligand that binds to a receptor and decreases response below basal level

Assays

I. Saturation Experiment: Measures binding equilibrium by performing a titration of radioligand, keeping the amount of receptor constant.

You can use saturation curves to determine Bmax (receptor expression level) and Kd (binding affinity of ligand).

saturation_experiment_ASK.jpg
Saturation experiment

  • Total binding: Increasing concentration of radioligand in absence of cold ligand. Measures both specific binding to receptor, as well as non-specific binding.
  • Non-specific binding: Increasing concentration of radioligand in presence of cold ligand. Measures binding of the radioligand to non-receptor components.
  • Specific binding: Total minus non-specific binding. Measures binding to the receptor, specifically.

II. Competition Experiment: Measures equilibrium binding of a single concentration of radioligand in the presence of various concentrations of unlabeled competitor.

You can use competition curves to determine the affinity of a large number of unlabeled compounds for a receptor in screening.  IC50 and Ki can be derived from competition data.

competition_ligand_ASK.jpg
Competition curve

  • IC50: Concentration of a competing ligand that displaces half of the radioactive ligand
  • Ki: affinity of a cold ligand for the receptor

Note: When changing concentration of radioligand, a shift in IC50 of a reference compound can be observed; however, the Ki stays the same.

III. Kinetic experiments: Measure the level of binding at various times to determine the rate constants for radioligand association and dissociation.

association_dissociation_ligand_ASK.jpg
Kinetics

  • Saturating concentration of ligand added an amount bound over time to determine the Kon
  • High concentration of unlabeled competitor added and decrease in binding over time measured to determine the Koff
Kd = K(off)
K(on)

Top

Tips and FAQs


Ideally, you will want to select a radioligand that has the following properties:

  • High specific activity (> 20 Ci/mmol for a tritiated ligand)
  • Low non-specific binding (hydrophobic ligands will generally show higher non-specific binding)
  • High purity (typically > 90% pure)
  • High selectivity
  • Stability (if you will need to use your radioligand over an extended period of time, stability may be a factor. 125I-labeled ligands should generally be used within one to two months of manufacture date. Tritiated ligands should usually be used within 3-6 months of manufacture date (however, there are exceptions to this).

You can view our GPCR/radioligand selector guide for assistance in choosing a radioligand for GPCR studies.

When plotting your data, we recommend that you use the experimentally-determined concentration of your radioligand working solutions, rather than a calculated concentration. Some radioligands will stick to the walls of your dilution tubes. It is important to determine the true concentration of your pipetted radioligand. To experimentally determine the concentration of a radioligand dilution, pipette a few microliters out of your dilution tube and spot onto a filtermat. You may also want to spot a 1/10 dilution of this sample, so that you can cross-check your results. Determine the dpms (disintegrations per minute). Convert from dpms to Curies using the following equation:

1 Ci = 2.22 x 1012 dpm

Then use the specific activity of your lot of radioligand to convert from Curies to moles. To determine the radioactive concentration at this point, divide the number of moles by the volume you spotted onto the filtermat.

FAQs

Q. What concentration of radioligand should I use for my saturation curve?
A. We generally recommend you choose 3-5 concentrations below the Kd, and 3-5 concentrations above the Kd. The highest concentration should be ten times the Kd.

Q.What concentration of radioligand should be selected for competitive binding?
A. The radioligand is used at a low concentration, usually at or below its Kd value. If the specific activity is low, concentration above Kd value can be used but it must never be at or higher than saturating concentration.

Q. What type of ligand should I use to determine non-specific binding in my saturation curves?
A. You usually want to choose a ligand that is different than your radioligand. You want to displace only the specific binding of the radioligand to the receptor. You should choose a ligand that has high affinity for the receptor binding site and low affinity for non-specific binding sites.