Water Analysis Using LAMBDA UV/Visible Spectrophotometers Hexavalent Chromium Determination | 爱游戏平台注册登录 - 爱游戏登录网,爱游戏平台注册登录
爱游戏平台注册登录

關於Revvity-- 以科技之能,突破人類潛能的邊界

(Revvity的前身是隸屬於爱游戏登录网 )


兩家獨立企業由此誕生

生命科學與診斷 & 分析和企業服務

Brief PMS299 D

Application Note

Water Analysis Using LAMBDA UV/Visible Spectrophotometers Hexavalent Chromium Determination

WaterAnalysisforHexavalent_512x288.jpg

Introduction

Chromium (VI) ions react with diphenylcarbazide in weakly phosphoric solution. The product, diphenylcarbazone, is a red-violet complex which can be detected photometrically at 550 nm. The chromate cell test kit is suitable for the concentration range of 0.11 – 4.46 mg/L chromate allowing the concentration of chromium(VI) in a water sample to be determined without the use of a calibration curve by multiplying the measured absorbance at 550 nm with a known factor.